A Digital HPLC Flowmeter Makes System Calibration Easier

Meeting tough FDA standards for HPLC system IQfree-standing and stack system options are the most
OQ PQ validation as well as ongoing PV (performancecommonly used although in some instances an
verification) calls for properly installed, calibrated, andadapter kit can be employed to mount the unit on
maintained equipment. A digital liquid hplc flowmeterwide necked reagent bottles.
such as the FlowCal 5000 is proven to be aWhile GLP requires annual recalibration, flow meter
convenient and accurate tool to calibrate HPLC pumpaccuracy can quickly deteriorate with improper use.
flow rates, simplifying the calibration and validationFor example, dirty sensing tubes can trap bubbles
procedures. Pump flow rates from 25 mL/min downand create false readings. In cases where bubbles are
to as low as 0.05 ml/min can be measured with anobserved during operation the quick remedy is turning
accuracy of ± 1 % of the flow rate. This type ofthe meter off and letting it drain, then return it to
tool saves time and labor compared with using aoperation. This suggests that the number one rule is
timer while manually collecting the effluent forto keep the sensing tube clean using supplied
measurement by weight or volume. In addition, theequipment for this important function. Another good
data collected is independent of operator techniquepractice is to institute a standard operating procedure
and is therefore both more accurate and morespecifying that flow meters be cleaned every two
precise.weeks when they are in constant use. Flow meters
When checking into digital HPLC flow meterssuch as the FlowCal 5000 are shipped with a cleaning
laboratory technicians should determine if the modelskit and directions for carrying out this function, a
being considered will deliver accurate data at the flowprocedure that is usually completed within 15 to 30
rates typically used and have a traceable calibrationminutes. If flow meters are not to be used over a
certificate. In addition, the flow meter must beweek or more they should be flushed with ethanol to
compatible with all brands of HPLC systems used inprevent algae growth.
the lab. All wetted parts should be able to safelyRecord Keeping
handle common HPLC solvents. The flow meterOne method to archive data in order to properly
should have an appropriate interface such as anrespond to FDA inquiries involves connecting the flow
RS232 port for connection to a printer or PC formeter to a PC using the RS-232 interface. This allows
printing and archiving in accordance with goodoperators to record flow meter readings
laboratory practices (GLP).electronically, such as in an Excel document using the
Installation and MaintenanceWindows HyperTerminal program. The process
Although they are highly accurate, digital flow metercollects data and displays it using HyperTerminl.exe,
performance is adversely affected by impropercopies it on a clipboard and pastes it into Excel for
mounting and poor maintenance. Citing the FlowCalanalysis. Step-by-step details for this process are
5000 as an example, several mounting configurationsavailable from FlowCal equipment providers such as
are available and supported by hardware suppliedTovatech.
with the unit. Regardless of the mounting method,To support labs that choose this approach, some
the unit must be within 10 degrees of vertical andHPLC flow meter suppliers offer an optional miniature
with suitable space available for placing the collectionthermal printer to record all readings and error
vessel below the flow meter. Wall mounting,messages output via the RS 232 interface.